Überblick

 
Plum 360 | Infusion System

Your Direct Connection to Clinical Excellence

With today’s hectic pace, tight budgets and patient safety concerns, you need an infusion pump that serves both you and your patients, helping to maximize valuable resources and enhance your clinical practice. You’ll find the Plum 360 infusion system is unlike any other.

  • Air Matters — the Plum 360 utilizes the unique PlumSet cassette that features air management that doesn’t require disconnecting from the patient

    • Remove air faster — use the pump to automatically remove air through the cassette’s secondary administration port without disconnecting the line from the patient

    • Reduce therapy interruptions — capture 1 mL of air in the cassette’s air trap – an industry exclusive feature – before the alarm is activated

    • Reduce risk of contamination — help eliminate the potential for contaminants by keeping the patient line connected

  • Secondary Delivery is of Primary Importance — ICU Medical's Plum infusion systems are unique in the way they connect the secondary line directly to the PlumSet cassette, so the pump can differentiate between primary and secondary infusion lines and deliver each as intended

    • No need to raise and lower IV containers — the Plum 360 infusion system functions independently of head-height

    • Simultaneously deliver two compatible medications at independent rates — through a single line – something only ICU Medical’s Plum infusion systems can do

    • Deliver medication as intended — the direct connect secondary delivery helps to ensure that every milliliter of medication is administered as intended, helping to improve delivery accuracy and avoid infusion errors

  • Optimize Safety and Compliance at the Point of Care — when programming your Plum 360 infusion system with ICU Medical MedNet software, you automatically start in the drug library

    • Large drug library — includes 2500 medications (medications used in multiple clinical care areas (CCAs) count as a single entry), 40 CCAs and 400 medications per CCA

    • Clinical use selection — multiple limits, based on clinical usage can be defined for a given medication, allowing further refinement of the infusion parameters

    • Fast drug search — using alphanumeric keypad to enter initial letters of the medication

  • EMR Ready, Today and Tomorrow — when you and your electronic medical record (EMR) partner are ready for integration, your Plum 360 infusion system with ICU Medical MedNet safety software is ready too. Whether it’s today, or any time after, ICU Medical’s IV Clinical Integration solution will connect your Plum 360 system to the medication order, patient information and EMR

    • Automated programming — automatically populate your Plum 360 pumps with medication orders sent directly from your pharmacy system

    • Automated documentation* — capture pump settings, start-and-stop times and titrations

    • Wireless integration — the expanded memory and 802.11 a/b/g/n wireless capabilities of the Plum 360 facilitates faster system-wide deployment of drug library updates, helping to improve safety

* Functionality dependent on EMR system capability

 

Plum 360™ Infusion System with ICU Medical MedNet 

 

Availability

In Stock

ICU Medical Product Numbers

List Number

30010

Order Quantity

1 Unit

Case Size

1 x 1

Pump Line

Plum 360

Pump Type

General Infusion

Dimensions

Approx. 8” H x 8” W x 6” D (0.20 m H x 0.20 m W x 0.15 m), excluding pole clamp extrusion and power cord storage

Weight

Approx. 10 lbs. (4.5 kg) with battery

Battery Life

Approx. 3 hours at up to 125 mL/hr on a single line, or delivers at least 250 mL if > 126 mL/hr

Power Requirements

120 V~, 50 to 60 Hz, 50 VA

Recharge Time

Approx. 8 hours with the device operating at 125 mL/hr on one line

Volume Range

0.1 to 99.9 mL (in 0.1 mL increments), 100 to 9999 mL (in 1 mL increments)

Delivery Rate Accuracy

± 5% (1 to 999 mL/hr), ± 10% (0.1 to 0.9 mL/hr)

Delivery Rates

Lines A and B: 0.1 to 99.9 mL/hr (in 0.1 mL increments), 100 to 999 mL/hr (in 1 mL increments)
Concurrent Delivery: 0.5 mL/hr minimum for each line
PlumSet: 500 mL/hr cumulative (A+B) maximum
KVO: 1.0 mL/hr or the last delivery rate on the associated line, whichever is less

Delivery Units

mL/hr, mcg/kg/min, mcg/min, mcg/kg/hr, mg/min, mg/m2, mg/m2/min, mg/m2/hr, mg/kg, mg/kg/hr, mg/hr, ng/kg/min, g/hr, mEq/hr, mUn/min, units/min, units/hr, units/kg/hr

System Alarms

Occlusion, Air-in-Line, Low Battery, Door Opened While Pumping, Lockout Violation, VTBI Complete, Valve/Cassette Test Failure, Nurse Call, No Action, Programming Not Complete

Occlusion Alarm Pressure

Distal Pressure Limit (without alarm): 1-15 psig (default 6 psig)

 